high temperature dry block calibrator Range 500 °C to 1200 °C

high temperature dry block calibrator Model ED-1200 is a best metrological product for calibration lab, highly recommended for high temperature sensors.

Features

  • HMI  Display.
  • Protection disc.
  • Great temperature stability.

Datasheet

 

Download_User_Manual

Description

Here technical specification of high temperature dry block calibrator

 

1. Temperature Range: 500 to 1200 ˚C.
2. Display resolution: 1 ˚C
3. Control accuracy: set point ± 0.5 ˚C .upto 1000 ˚C
: set point ± 1˚C .upto 1200 ˚C
4. Thermal stability:
± 0.5 ˚C @500 ˚C (calculated after stabilization time of 20 minutes).
± 1 ˚C @ 1200 ˚C (calculated after stabilization time of 20 minutes).
5. Thermal Nonuniformity (Radial) well to well:
± 2.5 ˚C @1200 ˚C Reference well 1.
6. Stabilization Time/settling time: 20 minutes after set point is achieved .
7. Time to reach ambient to 1200° C (40 Minutes).
8. Power supply: 240 VAC @50-60 Hz.
9. Current : 8 A.
10. Power Consumption: 2kWatt maximum.
11. Sound: 30 dB.
12. Enclosure Metal(SS) Powder coated(black)
13. Weight: 18.5 kg.
14. Dimensions: H:200, W:300, L:450mm.
15. Standard Insert 2* 8.2mm, 1*10.2mm.
16. Diameter Tolarance: ± 0.1mm

Performance of ED 1200 At 1200°C | Model No. - ED 1200 | Tunix Corporation

Performance of ED 1200 At 1000°C | Model No. - ED 1200 | Tunix Corporation

Performance of ED 1200 At 500°C | Model No. - ED 1200 | Tunix Corporation

Performance of ED 1200 At 1000°C | Model No. - ED 1200 | Tunix Corporation

Reviews

There are no reviews yet.

Be the first to review “high temperature dry block calibrator Range 500 °C to 1200 °C”